  • Lab Update - Run 78
    • Swapping DCRCs and cable seem to have helped in finding a config where all channels are responding to QET bias changes. On DCRC1, the QET bias change is minimal on Side 1 channels.
    • DCRC4 seem to sometimes respond to QET bias changes and sometimes not.
    • Tried using the ReVE board to look for traces with the new DAQ. Plugging in IO cable and checking pins on 100pin connector shows QET A, C and D but not B. The IO used has known bad QET C, so tested for QET A and D. Couldn't get to see traces with PulveViewer. Error related to not getting any new events.
  • Priority Updates