  • Lab Update - Run 78
    • Turning DCRCs ON raised the Tower temperature at 107 mK (from 76 mK).
    • SQUID tuning was done for DCRC1 and DCRC3. Tower temperature dropped to 99 mK.
    • DCRC4 is not showing feedback signal. Earlier, Matt reconnected the corresponding IO cable to DCRC1 for tuning and then set those values for DCRC4.
    • DCRC1 and DCRC4 have channels where trace baseline in not responding to QET bias changes to do IbIs. Will try to figure out what is causing that.
    • Haven't seen transition in any of the channels yet.
    • Ping is getting ready to handle LHe transfers since Zach will be leaving for Fermilab after next week.
  • Priority Updates
  • General Updates
    • Himangshu
      • Ziqing emailed about having a compressed air monitoring page for SNOLAB environmental variables setup this week before the CUTE run.