- Lab Update - Run 77
- Fridge still warming up, now at 250 K. But not warm enough (~ 277-280 K) to open it yet.
- Looked up historical thermal touch issues and saw a very similar one from this run diary entry
- Possible things to be look for: 1) Copper tape used to close holes can touch 100 mK stage to other stage (for example, 600 mK rod), 2) Crooked 100mK heat sinks can touch tower.
- For electrical shorts, 1) Redo kapton tape on the flyover cable meant to prevent cutting in and shorting the wires, but we could do this more carefully/thoroughly.