- Lab Update - Run 77
- Donot see phonon pulses after 4th 10k pumpout which was on Saturday
- After the 3rd 10K pumpout, we see a few channels on Side 1 in transition, while two of the channels were not transitioning as if they are stuck in normal mode.
- Side 1 pC has low feedback coil resistance but the reason for Side 1 pD is not known yet.
^ From IbIs with MC = 31 mK |||
| SQUET | Cable | DCRC channel | Detector channel | SQUID Bias | SQUID Lockpoint | After Third 10K pumpout | Previous Run |
|---|---|---|---|---|---|---|---|
| ZC-74 | 1 | DCRC1pC | Side1pC | -92.6 | -3.98 | no transition(feedback resistance low) | 235 |
| ZC-74 | 1 | DCRC1pD | Side1pF | -97.9 | -2.4 | 148 | 274 |
| ZC-31 | 2 | DCRC3pA | Side1pE | -98.1 | -6.12 | 135 | 246 |
| ZC-31 | 2 | DCRC3pB | Side1pB | -87.6 | -3.49 | 190 | 225 |
| ZC-31 | 2 | DCRC3pC | Side1pD | -44.4 | -2.86 | no transition | 258 |
| ZC-26 | 3 | DCRC4pC | Side1pA | -83.7 | -7 | 320 | 212 |
| ZC-74 | 1 | DCRC1pA | Side2pC | -95(guess) | -2.5(guess) | no transition | |
| ZC-74 | 1 | DCRC1pB | Side2pF | -100 | -2.01 | no transition | |
| ZC-31 | 2 | DCRC3pD | Side2pE | -89.1 | -3.81 | no transition | |
| ZC-26 | 3 | DCRC4pA | Side2pB | -83(guess) | -5.91(guess) | no response to change in QETbias | |
| ZC-26 | 3 | DCRC4pB | Side2pA | -83(guess) | -5.91(guess) | no response to change in QETbias | |
| ZC-26 | 3 | DCRC4pD | Side2pD | -93.1 | 1.56 | no transition |
* After fourth 10K pumpout, the working channels on Side1 are not showing pulses when we did IbIs on them.
- Tried using automated IbIs using pysequencer scripts. The sequence gets stuck at the step of Transition OFF (stopping the run).
Suggestion: odbedit
- Priority Updates
- General Updates
- Himangshu
- Shubham
- Jack
- Derek
- Zach
- Scott
- Giulia