  • Lab Update - Run 77
    • Got to base temperature friday, started seeing MC temp spikes just like last run.
    • Temperature in the Tower thermometers read 15-20mK higher than MC thermometer. New RoX temperatures are 10-20 mK higher than old tower temperatures.
    • Started MIDAS to look at the DCRCs with a 2-sided 0V setup. DCRC1 has all working channels, Two phonon channels in DCRC4 didn't show traces (pB and PD). No channels in DCRC3 showed traces.
    • Did another electrical check at BOB. On (DCRC1) Cable1 feedback C showed lower than expected resistance. On Cable 2(DCRC3) SQUID A showed high resistance. On Cable 3 (DCRC4) SQUID D showed high resistance.