- Lab Update - Run 76 ending
- NaI array is disassembled, lead/poly shield removed. Fridge warm.
- Can clearly see a leak at stripline elbow, nowhere else.
- tried pushing vacuum grease in up to the o-ring joint, didn't help
- will need to do a full disassembly
- last replaced the o-ring in May 2021, it was OK for over a year - problems first occurred in August 2022
- would a tighter o-ring be better?
- would fixing the stripline elbow in place be better? It kind of gets tilted by the e-box plumbing.