- Lab Update - Run 76 in progress
- NaI spectrum taken over a couple of days (one bin at a time)
- full-array spectrum of single-detector energies
- taken with PuBe present in standard position
- first NaI scan since the recalibration a while ago
- sharp peaks between 11 and 12 MeV
- these are saturation peaks. Each detector will have a saturation amplitude; all events at that energy or above will appear in one bin
- Noticed that the upper threshold I've been using is too high at 11.2 MeV - this includes some saturation peaks, so it wasn't cutting out muons
- Redoing NaI-coincident data now with upper threshold reduced to 11 MeV
- Reduced upper edge will do two things
- reduce trigger rate from around 11 Hz to around 8 Hz and thus reduce accidental background overall
- eliminate muon-related coincident events (at least for through-going muons)