  • Lab Update
    • Run 76 in progress
    • Base temperature better than last run: MC 27.8 mK, last run 29.8 mK
    • Goldman thermometer is reading wrong….
      • last run: Goldman thermometer (wired through striplines) and Tower thermometer (using fridge thermometer wires) - Goldman read correctly, Tower was wrong
      • between runs: checked for problems with fridge thermometer wiring. No leakage between wires seen at room temperature (> 1 GΩ)
      • this run: assumed Tower thermometer itself was the problem, so using Goldman thermometer with fridge wires
      • conclusion: something wrong with fridge thermometer wires.
      • Goldman reading ~ 80 mK but this is definitely not correct, probably ≤40 mK. Could be explained by a ~50 kΩ leak between wires… ?
      • Unclear how to fix this. Likely any new problem with these wires is somewhere near the bottom where there's activity. Maybe replace lower section of the wiring.
    • Half the detector channels showing good pulses using previous settings. Others need troubleshooting.
    • NaI array installed and working
    • Denver crew working on Run 76 DAQ