  • Lab Update
    • With Side1-readout HV board, best noise width achieved was 6.8 eVee (@ 100V with prebias, with dead pF but with source situated above live pC)
    • Now testing Side2-readout HVboard, for the first time
      • modified to connect to existing LEDs
      • first connection: some channels fine, others noisy
      • improvements in noise by shorting together two of the isolated ground planes
      • DCRC1 still has some high high-frequency noise; this gets a little smaller if HV cable disconnected
      • channels looked good enough to take some 100V data today
      • moving source from y=30mm back to center, will take some more data
      • need to modify CDMSbats to process for this setup
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Jack
      • LF noise update
    • Derek
    • Zach
    • Michael
    • Scott
    • Giulia - ibis_data.xlsx
    • Giovanna
    • Nikki
    • Pramod
  • Fun Update
    • Canceled