- Lab Update
- With Side1-readout HV board, best noise width achieved was 6.8 eVee (@ 100V with prebias, with dead pF but with source situated above live pC)
- Now testing Side2-readout HVboard, for the first time
- modified to connect to existing LEDs
- first connection: some channels fine, others noisy
- improvements in noise by shorting together two of the isolated ground planes
- DCRC1 still has some high high-frequency noise; this gets a little smaller if HV cable disconnected
- channels looked good enough to take some 100V data today
- moving source from y=30mm back to center, will take some more data
- need to modify CDMSbats to process for this setup