  • Lab Update:
    • Realized that the non-functional channel, Side1pF, was a difficult wirebond repair - had to do surgery to turn a QET into a hopscotch pad.
    • Checked out the Side1-readout HV board: we reloaded a lot of resistors for biasing Side2 but two of them need to be redone. Otherwise ready to try.
    • Checked out the Side2-readout HV board: this one has never been tested. Looks like there may be errors in 4 connections.
      • The Side2-readout board may not work as well - several channels cross ground planes (in the Side1-readout board only one does this)
      • If we can't repair Side1pF, we should be able to rewire the Tower so that we can use the Side1-readout board to read out detector side 2
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Jack
      • Progress on LF noise
      • Have rough LF template - look at prepulse region and define relative phase to the template