  • Lab Update: Continuing with 2-sided 0-V data at a few different positions, testing stability. Calculated that the sensitivity of gain (dP/dI) to temperature is around 5% per mK, though it varies a lot by channel (1% to 13%). I've been correcting QET biases for temperature variations to (hopefully) eliminate the gain variations.