  • Lab Update: Run 75 in progress
    • Changed settings on DCRCs which resulted in lower base temperature, 42 mK instead of 48 mK. Turning off 15V power should, presumably, depower the FETs and allow a lower base temperature. But turning off 15V power on DCRC1 causes massive heating (0.7 K). Turns out though that leaving 15V on on DCRC1 and turning it off on DCRCs 3&4 caused the fridge to cool.
    • Took another set of automated IbIs at new lower temperature
    • Continuing to move source (in 10 mm increments) and take data.
  • Priority Updates
    • Spending money in the budget. We have leftover money (because we haven't spent money on travel, for one thing) and should spend it before around April. Ideas:
      • Replace our computer cluster. The machines cdms1,2,3 and cdms-04-1,2,3,4 are considered outdated by IT and they are going to stop managing, turning all of those maintenance tasks over to us. Alternatively we could buy a new cluster that IT would manage. We'll meet with Chad today.
      • Tower-less detector mounting setup. We can probably achieve colder SQUETs and therefore better noise if we mount the detector without a Soudan tower. (For phonon-only devices - for charge readout we'd still need the tower.) Mostly need to design this, and can get ideas from e.g. TAMU. Need to buy superconducting wire and some sort of connectors, build plates and things (could have shop do the fabricating).
        • Prisca has asked whether a SNOLAB tower would be useful for us. I'm not sure.
      • Buy a new calibrated thermometer. Something seems wrong with Tower thermometer, and new Goldman thermometer is a little inconvenient to mount.
      • Laser and fiber-optic feedthrough for operating HVeV detectors. Could this also be used for low-energy calibration of 100mm Si detectors? Mostly need to figure out how to do this, but also will need to buy some things.
      • Service our spare circulation pump maybe.
  • General Updates
    • Himangshu
    • Jack
      • PCA analysis definitely has useful information, but maybe too much since we need 5-10 components to describe the desired features
      • Next step is to try to identify LF features from averaged traces, do more traditionally template analysis
    • Derek
      • Working on data that is on SLAC using SDF. Early stages right now, starting to get used to handling our data using python and sdf.
    • Zach
    • Michael
    • Scott
    • Giulia
    • Giovanna
    • Pramod