- Lab Update
- Data taken up to +75V and -75V. Continuing trend of higher drift rate needed at higher bias, steeper slope in -bias
- Luke gain looks pretty similar in - and + polarities so far - not yet seeing the strong polarity dependence noticed in previous runs
- Temperature rises several mK after LHe transfers and LN2 fills, then slowly settles, making it hard to take consistent data
- Studying whether to add more 4He to mixture