Batch mode: define a dataset, process it, read it back

A dataset definition is one small text file that says which raw files, how to read them and which processing to run. goldsluice runs it over every raw file in one command and writes the results next to it; goldpan opens the same file and gives you the pulses together with the processed quantities. Nothing is assumed: the readout, the channels and every parameter are written in the file. This page is the whole loop; the file format is specified in the docstring of goldpan.dataset (in the goldpan documentation), and a new algorithm is a different page (Running goldsluice and developing an algorithm).

You need the setup of that page (maturin develop --release in a clone of the goldsluice repository, or just pip install "goldsluice[batch]") and goldpan installed with Parquet support (pip install "goldpan[parquet]"; requirements.txt of NSDF-Data already does).

1. The definition file

A TOML file, by convention <name>.dataset.toml. Annotated; the whole file is examples/r76.dataset.toml:

[dataset]
name = "r76-07221203"
output = "~/goldsluice-output/r76" # the processed quantities: outside the repository

[[sources]]                        # where the pulses come from; several allowed
name = "side2"
format = "idx"                     # any goldpan adapter: idx, midas, parquet, ...
glob = "~/idx/07221203_2025_F*"    # raw files or folders: paths = [..] and/or glob
detector = "NAI-S2"
readout = "DCRC_RUN76"             # a goldpan.config preset, or three numbers
[sources.read]                     # the adapter's other read() arguments
channels = { Side2pC = [0, 0], Side2pF = [0, 1] }
roles = { Side2pC = "phonon", Side2pF = "phonon" }

[[processing]]                     # the steps, in order
algorithm = "peak"
channels = "role:phonon"           # "all", "role:<role>", or a list of names
[processing.params]
glitch_samples = "@readout.glitch_samples"       # from the readout above
pretrigger_samples = "@readout.pretrigger_samples"
amps_per_adc = 1.5e-9              # you must supply it: there is no default

[[processing]]                     # the next step, run in order
algorithm = "max_time"             # time of the largest sample, in s
channels = "role:phonon"
[processing.params]
sample_period_s = "@readout.sample_period_s"   # from the readout

Key

Rule

readout

A preset name or { pretrigger_samples = ..., glitch_samples = ..., sample_period_s = ... }, all three. Only needed by adapters that ask for it (idx, midas); a parquet source has its readout inside the file.

[sources.read]

Exactly the keyword arguments of that adapter’s read (help(goldpan.adapters.idx.read)); an argument it does not have is an error.

"@readout.<field>"

In a parameter: the value of pretrigger_samples, glitch_samples or sample_period_s of the data’s readout, so the file cannot disagree with itself.

Unknown keys

Errors, not ignored: a typo (readuot) fails at once instead of silently doing nothing.

2. Check, then run

python -m goldsluice algorithms
python -m goldsluice check r76.dataset.toml
python -m goldsluice run r76.dataset.toml

algorithms lists what can be named in algorithm = and the outputs of each. check reads the definition and lists the raw files it finds; it fails for an unknown algorithm, a pattern that matches nothing or a bad readout, before anything is processed. run processes the files and ends with a report that times each phase:

processed  .../raw/TOY/S1_F0001.parquet: 3000 events in 0.75 s
processed  .../raw/TOY/S1_F0002.parquet: 3000 events in 0.65 s
dataset big: 2 file(s) processed, 0 unchanged, 6000 events in 1.49 s
  read        1.244 s   (the adapters)
  compute     0.052 s   (peak)
  write       0.101 s
       4021 events/s overall

That is 6000 events of 3 channels and 4096 samples from a Parquet source. Reading is usually most of the time; the Rust kernels the least. Options: --chunk N (events read at once, default 1000: it bounds the memory used by the pulses), --force (redo files that are unchanged) and --backend python (the Python mirrors: slow, to compare numbers).

Results do not belong in the repository. check and run refuse an output folder that lies inside a Git working tree whose .gitignore does not cover the data files written there, and say so before anything is written. The repository ignores processed/ and _goldsluice.json, and the example writes to ~/goldsluice-output/, so the common cases pass. If you really mean it, add --allow-output-in-repo.

One output root. Set the environment variable GOLDSLUICE_OUTPUT to a folder (for example C:/NSDF/goldsluice-output) and a relative output in a definition (output = "r76-nai") goes under it, so the file holds no path of your machine and runs unchanged on another computer. A full path (output = "~/goldsluice-output/r76") is used as written; with the variable unset a relative output is relative to the definition file, as before. The goldpan writers (parquet.write, rq_root.write) write exactly where they are told and do not check the folder: the guard is batch mode’s.

Steps that build on each other. A step reads the raw trace unless it has an input. An algorithm that gives a trace instead of numbers (a filter or a baseline subtraction: python -m goldsluice algorithms says which: lowpass, subtract_line) needs a name and writes no quantity; a later step with input = "<name>" reads its result instead of the raw trace. A suffix is added to the names of the quantities of a step, so the same algorithm can run on both (adc_max and adc_max_lp). An input is a step above the one that reads it, and its channels must include the channels of the reader. The example examples/r76-registry.dataset.toml has two chains:

[[processing]]
algorithm = "lowpass"              # gives a trace: it has a name
name = "lp"
channels = "role:phonon"
[processing.params]
cutoff_hz = 95e3
order = 4
sample_period_s = "@readout.sample_period_s"

[[processing]]
algorithm = "peak"                 # reads the filtered trace
channels = "role:phonon"
input = "lp"
suffix = "_lp"                     # adc_max_lp, current_max_lp
[processing.params]
# glitch_samples, pretrigger_samples, amps_per_adc ...

Several sources, and different steps for each. A definition can have several [[sources]], each with its own files, format and detector (from the registry), so raw data from detectors that are wired differently go through one definition. Every step runs on every source, and channels = "role:phonon" means each detector’s own phonon channels. When the detectors need different processing, a step names the sources it runs on: sources = ["ge"]. A step without sources runs on all of them. A step that reads a filter (input) must run on a subset of the filter’s sources, and every source needs at least one step. A source’s files are redone only when the steps that run on it change. The sources’ different quantities read back together as records with different fields, so pick a source’s rows by series first. output_format = "root" is for one detector only.

Run metadata. Facts about whole series (high voltage, trigger setting) that an analysis needs for every event go in a small CSV, one row per series, with a series column and any others (hv_v, trigger_value, …; see examples/r76.series.csv). The definition only references it: metadata = "r76.series.csv" under [dataset] (relative to the definition file; later it may name a database instead). The values are joined when the data are read: Dataset.processed() adds one column per column of the table to every event, so a change of a value never redoes a file, and goldsluice run does not write them. A series with events that is not in the table is an error that names it; so is a column named like a quantity or a channel; processed(metadata=False) leaves them out. For events you already have, goldpan.runmeta.per_event(series, table) gives one value per event and groups(series, table, by=("hv_v", "trigger_value")) the events of each chunk.

3. Read it back with goldpan

from goldpan.dataset import Dataset

ds = Dataset.open("r76.dataset.toml")
ev = ds.read()         # the pulses with the processed quantities: one Events
q = ds.processed()     # only the processed quantities, no pulses: small, fast
ev.quantities["Side2pC"]["current_max"]    # a record per channel, a field per output

ds.iter_events() gives one Events per raw file (for a dataset larger than memory); ds.read(processed=False) gives only the pulses. Both packages read the same file, so a notebook and a batch job cannot disagree about what the dataset is. Reading before the batch run raises an error that names the command to run.

4. What is written, and what makes it run again

~/goldsluice-output/r76/
    _goldpan.json        the goldpan Parquet manifest
    _goldsluice.json     the ledger: per raw file, what was done, with what, how long
    NAI-S2/07221203_2025_F0001.parquet    one file per series and dump:
                         the key (series, dump, event) and the quantities

Any other tool that reads Parquet (pandas, Polars, DuckDB) opens these files. Each carries its provenance (the steps and parameters, the goldsluice version, the timing) in its metadata under goldsluice. A raw file is processed again when its size or date changes, or when anything that determines its results changes: the way it is read, the readout, a step or a parameter. Renaming the dataset or moving the output does not count. To start over, delete the output folder or run with --force.

5. Adding steps

Add another [[processing]] table. Each quantity must come from one step (two steps giving Side2pC.adc_max is an error that names it); different algorithms add different fields to the same channel’s record. Use channels = ["Side2pC"] to process only some channels. For an algorithm that does not exist yet, write its Python mirror and Rust kernel (the algorithm page): algorithms lists it as soon as it is registered.

6. ROOT files instead of Parquet

Two lines in [dataset] write the quantities as ROOT files laid out like the collaboration’s RQ files (they need uproot: pip install goldsluice[root]):

[dataset]
name = "r76-07221203"
output = "~/goldsluice-output/r76-root"
output_format = "root"             # "parquet" is the default
root_detector = 1                  # N of the zip<N> tree: required, never assumed

There is one file per series and dump, <series>_<dump>.root, with rqDir/eventTree (SeriesNumber, DumpNumber, EventNumber) and rqDir/zip<N> row-aligned, every branch a double, as in the real RQ files. The branches keep our names, Side2pC_adc_max and not the collaboration’s PAS1max: ROOT is only the container. The provenance of the run is stored as the string infoDir/goldsluice. The ledger, the chunks and the output-location guard work as for Parquet.

Read the files back with the RQ reader of goldpan (Dataset.processed() reads Parquet output only, and says so):

from pathlib import Path
from goldpan.adapters import rq_root

t = rq_root.read(Path.home() / "goldsluice-output" / "r76-root", detector=1)
t["Side2pC_adc_max"]        # one record per event

root_detector = 1 is our choice for the NaI detector of Run 76: the number is not in the raw data, so it lives in the definition file; use the same one in every definition of that detector. (The collaboration’s Z7 files have zip7 and zip14.) The commented lines in examples/r76.dataset.toml show the same two keys.

7. When it fails

Message

Meaning and fix

unknown key(s) [...]

A typo or a key that does not exist; the message lists the allowed ones.

...matches nothing

A paths or glob entry found no file; the message says where it looked (relative paths start at the definition file).

readout needs [...]

An inline readout lacks numbers; there are no defaults.

needs a readout

The adapter (idx, midas) takes a readout and the source gives none.

has no argument [...]

[sources.read] names something the adapter’s read does not take.

unknown algorithm

Not in algorithms; check the spelling or register the algorithm.

missing ... keyword-only argument

A step lacks a parameter: algorithms have no defaults.

nothing processed yet

You read the dataset before run.

has the batch run covered this file

New raw files were added after the last run; run again.

inside the Git working tree

The output folder is in a repository that does not ignore it: write it elsewhere (~/goldsluice-output/...), ignore it, or pass --allow-output-in-repo.

needs root_detector

output_format = "root" without the number of the zip<N> tree.

rq_root.read (from processed())

The output is ROOT: read it with rq_root.read, not Dataset.processed().

The numbers in current_max are in ADC counts times amps_per_adc: a placeholder of 1.0, as in the example file, gives counts.